|| TYPICAL ROOM LAYOUT
RadPRO® FM1
Floor Mounted Radiography System
Conventional Room
     
     
    View Next Layout
     
     

 

|| EQUIPMENT SCHEDULE
|| GENERAL NOTES
 
Equipment Identifier
 
BTU / HR
WEIGHT (lbs.)
   
Description
DIMENSIONS (in)
W x D x H
 
 
1701
RADPRO FM1
VSG-635 SERIES HF GENERATOR
CONTROL CONSOLE
17.05
11.73
40.28
75
 
RADPRO FM1
VSG-635 SERIES HF GENERATOR
LARGE POWER CABINET
23.30
14.17
27.16
165
 
250
RADPRO FM1
VSG-ET ELEVATING 4-WAY
FLOAT TABLE
86.60
31.50
35.40
656
 
1200
RADPRO FM1
VST-TS FLOOR-MOUNTED TUBESTAND
108.60
48.00
93.30
760
 
35
RADPRO FM1
VSG-WB VERTICAL CHEST STAND w/ BUCKY
23.10
13.40
88.00
319
         
         

All specifications and space requirements are shown generically. Contact virtual imaging for specific room requirements pertaining to optional equipment components and additional room preparation specifications

Absolute minimum room dimensions:
14’-0” W x 12’-0”

Minimum recommended ceiling height: 8’-0”

Suggested ambient room requirements:
66 to 75 degrees F at 30% to 60% r.m.
(Non-condensing).

Power Option #1
Three phase power (recommended)

Main power to generator:
280 vac to 480 vac, 3-phase, 60kva

Auxiliary power:
208/120 vac, 30 amp load center

Power Option #2
Single phase power (recommended)

Main power to generator:
240 vac, 1-phase, 60kva

Auxiliary power:
208/120 vac, 30 amp load center

Power Option #3
Single phase power

Main power to generator:
208 vac, 1-phase, 60kva
(with buck boost transformer)

Auxiliary power:
208/120 vac, 30 amp load center

 

© Planning Division. Virtual Imaging, Inc.

 
 
 
 